Coatronm2
Coagulation Analyzer
two channels, quarter volume
Cat.No. 80 700 00
Reduced reagent and sample volume with same precise and
accurate measurements make the routine more cost
effective.
This new manual Coagulation analyzer measure all
clotting based Coagulation tests with a quarter of the
regulary volumes.
Example:
PT: 25µl sample + 50 µl PT-reagent |
The
Coefficient of Variation (CV) is <2-5% depending on test
and reagent used.
The Coefficient of Variation (CV) is <2-5% depending on
test and reagent used. A new double cuvette allows an
effective and quick work, even for more samples. A
unique automatic amplification is especially designed
for all new sensitive Thromboplastin reagents and
equalize high differences of turbidity from reagents and
samples. All specific test parameters are stored in the
system.
The Coatron M2 fits perfect into small and medium
Coagulation laboratories. For usage in Emergency
laboratories this instrument offers all possibilities.
Tests
PT, APTT, TT, Fibrinogen (Clauss and PT derived),
Factor II, V, VII, VIII, IX, X, XI, XII, Fletcher, VT
(Venom Time),
APCR, AT-III (clot), Protein C (clot),
Protein S (clot), Heparin, STAT (PT/PTT)
Selectable dimensions
seconds, %-Activity, Ratio, INR, Units, g/l, mg/dl
Technical Description
The Coatron M2 is a micro-computer controlled
Coagulation instrument with two channels and a
LC-Display. The tests will be selected through keys on a
foil keyboard. The dialog language is user friendly and
selectable in English and German. Other languages can be
provided optionally
The Main program offers following menus
- ANALYSIS
- SETUP-TEST
- SETUP-SYSTEM
- SERVICE
The measurement can be
started manually by pressing the OPTIC-key in the same
time when reagent is added or with a connected
Autopipette. The results are displayed in the measured
dimension and can be converted into other desired
dimensions by pressing the UNIT-key. A connected printer
will automatically print out the results in all
dimensions with Optic channel and test name. A patient
identification No. can be optionally selected between
"NO PAT.ID", "MANUAL ENTRY" and "AUTO-SERIES" with Start
No.
Using the TECO Softwaretools the measuring curve data
can be transmitted to a Personal Computer and loaded by
any spread sheet with graphic function. Software updates
are realized with an easy download from a computer to
the instrument.
Technical Data |
|
Device |
SMD Board, V-25
Microcomputer, 18-bit Analog-Digital Converter, 1 MBit
Flash-EPROM; 1 MBit E2 PROM, |
Power Supply
(external) |
Input voltage: 94 to 243
Vac, Input frequency: 50 to 60 Hz, Power output (max):
40 VA,
Approvals: CSA, UL, TÜV-GS, CE-Mark |
|
Optic block |
2-channel Optics |
|
Keyboard |
Foil keyboard, 3x8 matrix,
with Test-, Functions- and Numerical keys |
|
Display |
4 lines x 20 Characters,
Liquid Crystal Display (LCD) |
|
Incubation block |
12 prewarming positions, 2
measuring positions,
1 Reagent position, mixed with different adapters,
1 Reagent position Ø 16 mm,
1 Reagent position Ø 22,5mm,
Temperature: 37,0°C ± 0,2°C |
Autopipette:
(optional) |
4-stage Autopipette with
electronical start
Volume: 25/50/100/200 µl |
Thermal Printer:
(optional) |
external, serial printer
with print-out of all results, print-out of test
parameters, print-out of system parameters |
Standard package
- 1 Pc Coatron M2
- 1 Pc Power Supply
- 1 Pc Mains cable
- 25 Pcs Double
cuvettes
- 5 Pcs Reagent
container
- 1 Pc Magnetic
stirrer
- 1 Pc Reagent
adaptor
- 1 Pc Printer cable
- 1 Pc Pipette
fastener
- 1 Pc Operation
Manual
- 1 Pc Warranty card
- 1 Pc Installation
report
- 1 Pc Service report
|
Consumables and
Accessory
- 19 000 02 Double
cuvettes
- 19 029 00 4-stage
Autopipette
- 19 030 01
Reag.adaptor Ø 22,5mm
- 19 030 02
Reag.adaptor Ø 22,8mm
- 19 030 03
Reag.adaptor Ø 24,2mm
- 19 030 04
Reag.adaptor Ø 27,8mm
- 19 030 05
Reag.adaptor Ø 25,2mm
- 80 050 06 Stirring
magnets
- 80 525 00 Reagent
container
- 80 530 00 Reagent
tubes
- 80 535 00 Thermal
printer
- 80 920 07 Pipette
tips
- 80 920 12 Thermal
paper
- 80 920 14 Printer
cable
|
|
Reagents |
- TEClot
reagents (PT, APTT, TT, FIB, VT, LA, PC, PS)
- TECal Normal
and Abnormal
- TEControl N,
A1, A2, LA+, PC, PS etc.
- Deficient
Plasmas
|
Important advantages
- all clotting tests
in one system
- quarter volume for
less price/test
- Research
capabilities
|

|
|
|
|